Maintenance
19
Drain off remaining operating fluid.
– In case of serious contamination, the operating fluid will have to be changed sev-
eral times (flushing):
Flushing
Fill up with operating fluid to the middle of the sight glass.
Operate the pump with the gas ballast open until the pump has warmed up.
Drain the operating fluid again and check for contamination, flush again if necessary.
Screw the operating fluid drain screw back in.
Fill up with operating fluid and check the filling level
Request safety data sheets for operating fluids and lubricants
from Pfeiffer Vacuum or download them from the Internet.
Dispose of operating fluid according to the local regulations.