Installation
11
Lay piping from the pump sloping downward so that no condensate can flow back into
the pump; otherwise fit a condensate separator.
– If an air trap is created in the system, then a device for draining condensation water
must be provided at the lowest point.
5.5 Connecting to the mains power supply
Single phase motors
The vacuum pumps are fitted with a built-in thermal protection switch. Depending on the
type of pump, different motor versions with a power cord are possible:
Fig. 6:
Motor circuit diagram with switch
WARNING
Emission of toxic substances from the exhaust!
Danger of poisoning from emitted gases or vapours, which can be detrimental to health
and/or can pollute the environment, depending on the particular application.
Comply with the applicable regulations when working with toxic substances.
Only officially approved filter systems may be used to separate and remove these
substances.
NOTICE
Excess voltage!
Danger of destroying the motor.
Power connections must comply with local regulations. Voltage and frequency infor-
mation given on the motor rating plate must correspond to the mains voltage and fre-
quency values.
To protect the motor and supply cable in case of malfunction, mains fuse protection
must be implemented.
WARNING
Danger of injury from moving parts!
After power failure or motor shutdown due to overheating, the motor may restart auto-
matically.
Secure the motor so that it cannot be switched on while any work is being performed
on the pump.
If necessary, dismantle the pump from the installation for inspection.