16
Maintenance
6.4 Switching off the pump
The pump can be switched off in any pressure range.
Rotary vane pumps have an integrated safety valve on the intake side. If the differential
pressure between the exhaust side and the intake side is
≥
250 hPa, then the valve clos-
es automatically and vents the pump when the pump is switched off.
Switch the pump off at the mains switch or disconnect from the mains in a secure man-
ner.
Venting the vacuum
chamber
Maintaining the vac-
uum in the chamber
7
Maintenance
7.1 Precautions
NOTICE
Danger of backflow of operating fluid into the intake line!
Contamination of the connected vacuum system!
Vent the vacuum chamber within 30 s, regardless of the chamber size.
For a longer venting process, use an additional shut-off valve and shut off the intake
line after switching off the pump.
NOTICE
Danger of backflow of operating fluid into the intake line!
Contamination of the connected vacuum system!
Because the safety valve of the pump is not suitable for longer-term sealing, install
an additional shut-off valve in the intake line.
Shut off the intake line immediately after switching off the pump.
WARNING
Danger of injury from moving parts!
After power failure or motor shutdown due to overheating, the motor may restart auto-
matically.
Secure the motor so that it cannot be switched on while any work is being performed
on the pump.
If necessary, dismantle the pump from the installation for inspection.
WARNING
Pump parts may be contaminated from pumped media!
Danger of poisoning due to contact with harmful substances.
Decontaminate the pump before carrying out any maintenance work.
In the event of contamination, take suitable safety precautions to prevent your health
from being harmed by any dangerous substances.