M300 Corpus HD
6.3.5
Example of error messages and
remedies
Event
Indication
Remedy
1 LED
Low battery voltage
Check the condition of the batteries. Check the contact between the
battery and the control unit.
2 LEDs
Failure in left drive motor
Check the connection of the left drive motor.
3 LEDs
Short circuit in left drive
motor
Check the left drive motor’s contacts and cables.
4 LEDs
Failure in right drive
motor
Check the connection of the right drive motor.
5 LEDs
Short circuit in right drive
motor
Check the right drive motor’s contacts and cables.
6 LEDs
Battery charger connected
Disconnect the battery charger.
7 LEDs
Joystick error
Check that the joystick has not been moved when starting the wheelchair.