3.4
Brake Error
This occurs when the control system detects a problem in the solenoid brakes
or the connections to them.
1505 - M1 Brake Error.
1506 - M2 Brake Error.
•
Check the solenoid brakes, cables and connections to the control system.
If the trip is still present after the above checks have been made, then the Power
Module may be defective. Refer to Section 5.
3.5
Motor Error
This occurs when the control system detects that a motor has become dis-
connected.
3B00 - M1 Motor Error.
3C00 - M2 Motor Error.
•
Check the motors, cables and connections to the control system.
If the trip is still present after the above checks have been made, then the Power
Module may be defective. Refer to Section 5.
3.6
Inhibit Active
This occurs when any of the Inhibit inputs are active and in a latched state.
The actual inhibit that is active is indicated by the last 2 digits in the Trip Code.
The code is in Hex.
1E01 - For Inhibit 1.
1E09 - For Inhibit 9.
1E0A - For Inhibit 10.
•
Cycle the power. This will drop out of Latched Mode which might clear the trip.
•
Check all wiring and switches connected to the indicated Inhibits.