Owner´s Manual Permobil C500
R-net Diagnostics
88
3.19 Bad Cable
This occurs when the control system detects a fault in the wiring in the communi-
cation cables between any of the modules.
•
Check all cables and connections for continuity.
•
If there is any visible damage to cables, replace and cycle power.
•
Disconnect one cable from the system at a time cycling the power after each
disconnection.
If the trip is still present after the above checks have been made, then the PM may
be defective. Refer to Section 5.
3.20 Bad Settings
This occurs when the control system detects incorrect or invalid program settings.
•
Check all parameter settings and re-program the control system using the R-net
PC Programmer.
•
Make a note of the current parameter settings and then reset the control system
to default settings.
•
Re-program the required settings in small groups, cycling the power after each
group to see if the trip occurs.
If the trip is still present after the above checks have been made, then the PM may
be defective. Refer to Section 5.
3.21 Module Error
This occurs when the control system detects a trip within a specific module. The
module will be identified on the diagnostics screen as described in section 2.
•
Check all cables and connections.
•
Cycle the power.
If the trip is still present after the above checks have been made, then the module
identified may be defective. Refer to Section 5.