Owner´s Manual Permobil C500
R-net Diagnostics
87
3.17 Memory Error
This is a non specific memory error which could be caused by any of the modules
within the system.
•
Check all cables and connections.
•
Cycle the power.
If the trip is still present and the system contains 3rd party Modules:
•
Disconnect all the non PGDT modules and cycle the power.
If this has cleared the trip:
•
Connect each 3rd party module in turn, cycling the power each time.
•
If the trip reappears after one of the power cycles then the last module to
have been added to the system must be defective.
If the trip is still present after the above checks have been made, then the PM
may be defective. Refer to Section 5.
3.18 PM Memory Error
This is a specific Power Module based trip.
•
Check all cables and connections.
•
Using the R-net PC Programmer, re-program the control system.
This should be done with either the most current specific program file for the wheel-
chair or the Permobils original programming file.
If the trip is still present after the above checks have been made, then the PM may
be defective. Refer to Section 5.
NOTE
Programming should only be conducted by healthcare professionals with in-depth knowledge of
PGDT control systems. Incorrect programming could result in an unsafe set-up of a wheelchair for
a user. Permobil accept no responsibility for losses of any kind if the programming of the control
system is altered from the factory pre-set values.