206 . Torion T-9 User’s Guide
To add a peak to the target library or a library in the Library Editor position the cursor on a peak in
the TIC so the spectrum is visible in the mass spectrometer graph or any of the spectra charts listed
above to get a spectra that has the desired masses to be added.
Right click in the spectrum chart and select the
Add Peak to Library
option. The following image
shows the dialog box that is associated with the option.
Peak Details
This section of the dialog contains the following.
Name
- This is the name of the compound that will be added to the Target List or a library. It
can be called anything. It does not need to match the library name, if desired.
CAS
- This is the CAS Registry Number associated with a compound and is a unique numerical
identifier assigned by Chemical Abstracts Service. The CAS numbers for all compounds can be
found in the AMDISNISTEPACW library which ships by default with the Chromion software.
When clicking
Search Libraries…
a search window opens in which to enter a
compound name to find a possible CAS number for the compound. The library that is
searched is the AMDISNISTEPACW which is the default library in Library Editor.
Destination
- There are two destinations to which the compound can be copied.
1.
Add to current target list
- If this option is selected the compound will be added
to the target list of the currently loaded method.
2.
Add to library
- When this option is selected a drop down menu becomes available
from which a library other than the default library can be selected. These are
libraries that are obtainable through
Library Editor
. If a new library is desired it
would first need to be created in the Library Editor prior to adding the new
compounds.
Source
- This section contains the retention information. This field is affected by other options
and by the current target list settings. For the search algorithms to work correctly unknown
libraries cannot use retention time and must have a retention index value.
•
Use retention TIME
- This option is only available for target lists that are using retention
time. If the target list in the currently loaded method is using retention index values this
option will not be available.
Summary of Contents for TORION T-9
Page 1: ...TORION T 9 Gas Chromatograph Mass Spectrometer GC MS User s Guide GC MS...
Page 4: ...iv Torion T 9 User s Guide...
Page 10: ...6 Torion T 9 User s Guide...
Page 11: ...1 Introduction...
Page 15: ...2 Safety Practices...
Page 24: ...20 Torion T 9 User s Guide...
Page 31: ...3 System Overview...
Page 40: ...36 Torion T 9 User s Guide...
Page 41: ...4 Basic Custodion Operations...
Page 49: ...5 Basic Operation...
Page 116: ...112 Torion T 9 User s Guide...
Page 117: ...6 Software...
Page 257: ...7 Advanced Operations...
Page 260: ...256 Torion T 9 User s Guide...
Page 261: ...8 Maintenance...
Page 268: ...264 Torion T 9 User s Guide...
Page 269: ...9 Troubleshooting...