Target Library Editor . 157
Target List Setup Tab
The
Target List Setup
is a subset library used to target compounds of particular interest to you. It
is different from the unknown library because it contains specific information that is influenced by
some of the instrument settings.
Mass List
The
Mass List
table shows the
specific ions that
are associated with each compound in the target
list. The following image shows the masses for Acetone.
Normally the tallest mass peak is designated as the
Main
ion and has an
Intensity
of 100. Each of
the other masses has an
Intensity
that is some percentage of the parent peak. For example the
second mass of Acetone 58 has an expected intensity that is 32% of the parent peak. The
Confidence
is an absolute value. So the
Confidence
level for the 58 mass is 50. This allows the 58
mass to vary from 25% of the parent peak to 125% of the parent peak and still be acceptable.
There are three options for the
Type
column. They are
Main
,
Necessary,
and
Unnecessary
.
•
As described above there is only one
Main
mass peak and it is usually the largest peak in the
spectrum. It is always assigned the value of 100% and a
Confidence
of 0. The
Confidence
value is obtained by taking the ratio of a mass peak to the
Main
mass. As a result the
Main
mass peak does not need a
Confidence
level.
•
Necessary
masses must be present in a spectrum or the search algorithm will not identify
the peak.
•
Unnecessary
masses are in the spectrum but are not used by the search algorithm to
identify a compound.
Unnecessary masses are usually masses that are less than 20% of the main mass. The
search algorithm uses all of the masses to determine a goodness of fit and a peak area
during the library search process. That is why even though a mass may be considered
unnecessary for identification it is still important to leave these masses in the table.
The values in the masses table can be edited by highlighting the entry with the mouse and
typing in the new value.
Summary of Contents for TORION T-9
Page 1: ...TORION T 9 Gas Chromatograph Mass Spectrometer GC MS User s Guide GC MS...
Page 4: ...iv Torion T 9 User s Guide...
Page 10: ...6 Torion T 9 User s Guide...
Page 11: ...1 Introduction...
Page 15: ...2 Safety Practices...
Page 24: ...20 Torion T 9 User s Guide...
Page 31: ...3 System Overview...
Page 40: ...36 Torion T 9 User s Guide...
Page 41: ...4 Basic Custodion Operations...
Page 49: ...5 Basic Operation...
Page 116: ...112 Torion T 9 User s Guide...
Page 117: ...6 Software...
Page 257: ...7 Advanced Operations...
Page 260: ...256 Torion T 9 User s Guide...
Page 261: ...8 Maintenance...
Page 268: ...264 Torion T 9 User s Guide...
Page 269: ...9 Troubleshooting...