88 . TGA 4000 Installation and Hardware Guide
10.
Remove the thermocouple by disconnecting the thermocouple connector and gently
pulling the thermocouple out of the furnace.
11.
Install the new thermocouple and replace the sample holder, alumina furnace cover
disk, and the alsiflex strip.
12.
Replace the lid holder and furnace lid.
13.
Replace the base plate thermocouple cover.
14.
Place the autosampler on top of the base plate, aligning the screw holes.
15.
Attach the autosampler to the base plate with the four screws.
16.
Place the sample trays on the autosampler with tray 1–22 toward the front and tray
23–45 on the rear.
After installing a new sample thermocouple, you will have to perform the gripper alignment
procedure.