AS 6000 Autosampler . 87
Replace the Sample Thermocouple in a
TGA 4000, Pyris 6 TGA or STA 6000 with
AS 6000 Autosampler
1.
Switch off the instrument.
2.
Remove both sample trays.
3.
Remove the four screws that attach the autosampler to the base plate.
NOTE: Never remove the screws from the plate that is directly attached to the analyzer's
cover. This plate is positioned with respect to the furnace. The positioning is
performed by the service engineer upon installation.
4.
Place your fingers in the indents in the sides of the base plate and lift the autosampler
up and away from the instrument.
5.
Remove the furnace lid and the lid holder.
6.
Remove the base plate thermocouple cover.
7.
Remove the alumina furnace cover disk.
8.
Remove the alsiflex strip.
9.
Remove the sample holder.