Changing the Sample Thermocouple
The sample thermocouple has a limited lifetime because of the range of temperatures and
the high maximum temperature it withstands. Therefore, the TGA 4000 and Pyris 6 TGA are
designed so that a user can change the sample thermocouple.
To remove the sample thermocouple, perform the following procedure:
1.
Switch off the analyzer and the remove the power cord.
2.
Remove the furnace lid.
3.
Remove the porous ceramic ring.
4.
Remove the sample holder.
5.
Remove the three screws of the top stainless steel furnace plate and then remove the
plate.
6.
Lift the tape off the top of the furnace to free the thermocouple.
7.
Unplug the yellow thermocouple plug.