Install a TGA 4000 . 45
Level the TGA 4000
1.
Lift the analyzer up at the front and turn each foot until it is screwed all the way into
the bottom of the analyzer.
NOTE: The analyzer is heavy. When lifting it along the front edge, be sure your hand is
firmly underneath it and can support the analyzer while you adjust the feet.
2.
Position the analyzer in the exact spot on the bench where you will be using it.
3.
Check whether the analyzer rocks on the bench, or if one foot is not touching the
bench. If so, lift the analyzer and slightly turn the foot clockwise to extend it. Lower the
analyzer to check that it does not rock. Once all of the feet are touching the bench,
make sure that the analyzer is steady.
4.
Place the level provided in the Spares kit on the furnace plate on the top of the
analyzer.
5.
Look at the level to see if the bubble is centered. If not, readjust the feet as in step 3
until the bubble is centered in the level.
The TGA 4000 is level when the bubble is centered inside the ring.
6.
Once the analyzer appears to be level, again make sure all of the feet are in contact
with the bench and that the analyzer does not rock.
This procedure is complete.
Summary of Contents for TGA 4000
Page 1: ...TGA 4000 Installation and Hardware Guide THERMAL ANALYSIS ...
Page 5: ...Introduction ...
Page 7: ...Safety and Regulatory Information ...
Page 17: ...Prepare the Laboratory ...
Page 20: ...20 TGA 4000 Installation and Hardware Guide ...
Page 21: ...Install Multiple Analyzers ...
Page 24: ...24 TGA 4000 Installation and Hardware Guide ...
Page 25: ...Install a TGA 4000 ...
Page 54: ......
Page 55: ...TGA 4000 Hardware ...
Page 67: ...AS 6000 Autosampler ...