60 . Raman IdentiCheck Getting Started Guide
SOP-7: Wavelength Calibration of the Raman
IdentiCheck
Wavelength calibration should only be performed if the instrument has failed a wavelength
calibration verification. This procedure does not use the triggered fiber optic probe. It uses
an internal neon standard.
1.
Cover the tip of your triggered fiber optic probe with a lens cap or a thick, black cloth,
to ensure that no light can enter the probe.
2.
Ensure the IdentiCheck is switched on and is stable, by following SOP-1.
3.
Ensure that the Triggered Fiber Optic Probe is selected as the Accessory; see
Commissioning your Instrument
on page 48.
The Probe Path LED on the probe will be illuminated.
4.
Select the Setup Instrument Advanced tab and click
Toolbox
.
The Adjustments Toolbox dialog is displayed.
5.
Select
.
The Wavelength Calibration dialog is displayed.
6.
Click
Next
.
The instrument will now perform the wavelength calibration routine. This will take some
time. At the end of the calibration verification, the result is displayed.
The instrument is now wavelength calibrated.
Summary of Contents for Raman IdentiCheck
Page 1: ...Raman IdentiCheck Getting Started Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 Raman IdentiCheck Getting Started Guide ...
Page 11: ...Safety Practices ...
Page 26: ...26 Raman IdentiCheck Getting Started Guide ...
Page 27: ...Raman History Theory and Instrumentation ...
Page 34: ...34 Raman IdentiCheck Getting Started Guide ...
Page 35: ...Introduction to the Raman IdentiCheck ...
Page 50: ...50 Raman IdentiCheck Getting Started Guide ...
Page 51: ...Using the Raman IdentiCheck ...
Page 65: ...Maintenance and Troubleshooting ...
Page 73: ...Appendix A Using your own PC ...