Instrument Methods . 43
Multi Wavelength
The Multi Wavelength method can provide absorbance readings at up to 5 wavelengths.
The procedure is as follows:
1.
From the home page, press 1 to open the Standard Methods folder.
The Standard Methods screen is displayed.
2.
Press 2 to open the Multi Wavelength method.
The Multi Wavelength – Method Settings screen is displayed.
3.
Select the number of
Wavelengths
and then press the down arrow.
The options are between 2 and 5.
4.
Enter the first wavelength (
λ1
) using either the number keys or the left and right
arrows, and then press the down arrow.
5.
Repeat Step 4 for each of the wavelengths.