Routine maintenance
56
4.
Lift off the mirror module.
5.
In a similar way replace it with the mirror module you want. When the
mirror module is in place, press down on the sprung metal piece to
release the mirror module so that you can unhook the tool.
6.
Close the lid when you have finished.
7.
The instrument will read the barcodes on the mirror modules and update
the list of mirrors in the software.
Changing the bottom mirror
Note!
This operation has to be performed by touch and not by sight so the
pictures do not show the components when they are actually installed inside
the instrument.
1.
Release the white cover on the right side of the instrument by twisting it.
Remove it.