Routine maintenance
52
2.
Push the plate carrier gently inside the instrument into the right upper
corner of the xy track
3.
Remove the tip mount if installed
4.
Wet a microfiber cloth (or some soft non-dust tissue) with pure ethanol.
5.
Put your hand into the instrument through the plate door of the
instrument and clean the lens (see picture below, in which the horizontal
head is shown vertically). The lens is not directly visible but can be
cleaned from the outside. With a small mirror it is possible to check that
the lens look clean.
With plate turned around,
remove tip mount
Clean lens through plate door