Functional description
26
Figure 30. Laser excitation from above to adjacent well/reading from
above.
Both excitation and emission occur from the top of the sample. Bottom
reading is not possible with AlphaScreen in EnVision.
Only one measurement per well is recommended because the sample is
partially bleached by the excitation light.
Crosstalk is a factor that must also be taken into account. There are three
components: bleaching caused by the excitation of adjacent samples,
afterglow from a previously excited adjacent sample and glow from an
adjacent well at the time of measurement. A crosstalk correction wizard
allows the amount of the crosstalk to be measured and the software uses this
as a correction factor in the calculation of the final result.
TRF laser mode
Figure 31. Laser excitation (second light source) from above/reading from
above. Bias mirror module with a single mirror is used.
Summary of Contents for EnVision
Page 1: ...EnVision Multilabel Reader 2104 9030 02 March 2008 INSTRUMENT MANUAL ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 11: ...1 Chapter 1 Introduction ...
Page 12: ...2 ...
Page 15: ...5 Chapter 2 Functional description ...
Page 16: ...6 ...
Page 45: ...35 Chapter 3 Information about user instructions and warnings ...
Page 46: ...36 ...
Page 59: ...49 Chapter 4 Routine maintenance ...
Page 60: ...50 ...
Page 75: ...65 Chapter 5 Specifications ...
Page 76: ...66 ...
Page 86: ...Specifications 76 ...
Page 87: ...77 Chapter 6 Quality information ...
Page 88: ...78 ...
Page 90: ...80 ...
Page 92: ...82 ...
Page 93: ...83 Chapter 7 Glossary ...
Page 94: ...84 ...
Page 104: ......
Page 105: ...95 Chapter 8 Installation ...
Page 106: ...96 ...