68153 01
www.perkinelmer.com
15
XRpad2 4336
D
IGIT
AL
I
MAGING
•
Check to ensure the XRpad2 LBP is clean, dry, and free of foreign contamination or debris. If
cleaning is necessary, see
“Section 8.5.1.6, Cleaning the XRpad2 LBP” on page 17
for cleaning
instructions.
•
Ensure the XRpad LBC charger is powered on.
•
Orient the XRpad2 LBP to match the orientation of the XRpad LBC Charger, and insert the
XRpad2 LBP firmly into the XRpad LBC charger. Keep the XRpad LBP in the XRpad LBC
charger until all the four charge status LEDs maintains a solid green, indicating a full charge.
To remove, lift the battery out of the XRpad LBC charger.
8.5.1.2
Installing the XRpad2 LBP
When installing the XRpad2 LBP in the X-ray detector in place of the XRpad2 Protective Insert,
or to change a used XRpad2 LBP, remove the XRpad Protective Insert or used XRpad2 LBP
before installing the new XRpad2 LBP (see
“Section 8.5.1.3, Removing the XRpad2 LBP” on
To install the XRpad2 LBP:
1
Fully support the X-ray detector to prevent it from dropping or slipping.
2
Check to ensure the battery compartment of the X-ray detector is clean, dry, and free of
foreign contamination or debris. If cleaning is necessary, see
“Section 8.5.1.6, Cleaning the
for cleaning instructions.
3
Check to ensure the XRpad2 LBP
is clean, dry, and free of foreign contamination or debris. If
cleaning is necessary, refer to
“Section 8.5.1.6, Cleaning the XRpad2 LBP” on page 17
for
cleaning instructions.
4
The XRpad2 LBP is keyed and can only be inserted into the XRpad2 4336 detector in one ori-
entation. Align the orientation of the XRpad2 LBP to match the orientation required on the X-
ray detector.
5
Insert the charged XRpad2 LBP into the X-ray detector in the corresponding orientation, and
gently press on the end cap until the latches secure the XRpad2 LBP inside the detector.
WARNING
Do not drop or hit the XRpad2 LBP against hard objects as this may
cause a risk of damage to the XRpad2 LBP, which may result in
exposure to the corrosive cell contents, fire, or explosion.
WARNING
There is risk of explosion, personal injury, or damage to the product
if the Battery XRpad2 LBP is replaced by a non-OEM approved
component.