XRpad2 4336
16
68153 01
D
IGIT
A
L
I
MAGING
6
Push the
Power
button on the X-ray detector to power on.
7
Check the battery charge status on the X-ray detector. If the battery charge status shows
sufficient battery charge is present, the X-ray detector is ready for use. If the battery charge
status shows lower than desired battery charge level, replace the battery with a charged
battery.
8.5.1.3
Removing the XRpad2 LBP
To remove the XRpad2 LBP:
1
Fully support the X-ray detector and XRpad2 LBP before performing this task to prevent
them from dropping or slipping.
2
Power off the
X-ray detector by pressing the
Power
button on the X-ray detector.
3
Move the two sliding latches closer to the center to disengage the XRpad2 LBP from the X-
ray detector (see
). Remove the XRpad2 LBP from the battery compartment of the X-
ray detector using a slow and steady pulling motion, supporting both the X-ray detector and
the XRpad2 LBP.
4
Store the XRpad2 LBP in a cool, dry, clean environment if not in use or when recharging the
XRpad2 LBP for the next use.
Figure 5
Removing the XRpad2 LBP
8.5.1.4
Transportation and Storage
When transporting and storing the XRpad2 LBP:
•
Follow all local, state, and federal/national regulations for handling, packaging, labeling,
and transporting Lithium-Ion batteries.
•
Before transporting the battery, inspect it to confirm that there is no damage or leakage from
the battery.
•
When possible, transport the battery in a discharged state.
•
Transport the battery in approved packaging only. Retain the original packaging, inclusive of
the Safety Data Sheet (SDS), for transporting the battery.
•
Store the battery in a cool, dry, clean environment if not in use. Do not remove the battery
from its original packaging until required for use.
Caution
Dispose of a used XRpad2 LBP according to the instructions in
“Section 11.0, Disposing of the X-ray Detector” on page 35
.