8.2
Environmental Considerations
includes a list of environmental considerations for the transport, storage, and operation
of the X-ray detector.
WARNING
Storage or use of the X-ray detector and power supply in
environmental conditions outside the specification may cause fire,
electrical shock, and unknown hazards, which may result in severe
personal injury, death, or substantial product damage or reduced
product lifetime.
Table 6
Environmental Considerations
Environment
Transportation/Storage
a
a. In original transport container for 365 days.
Operation
Ambient Temperature
b
(30d/365d)
b. Temp. Gradient: max 4.5 K/hour.
–10°C to 55°C/0°C to 55°C
10°C to 35°C
Relative Humidity
5% to 90%
10% to 85%
Atmospheric Pressure
520 hPa to 1070 hPa
690 hPa to 1070 hPa
Vibration
c
(EN60068-2-64)
c. Image quality cannot be guaranteed during shock or vibrations.
5 m
2
/s
3
(10 Hz to 100 Hz)
1 m
2
/s
3
(100 Hz to 2000 Hz)
0.5 m
2
/s
3
(10 Hz to 100 Hz)
0.1 m
2
/s
3
(100 Hz to 2000 Hz)
Shock
c
(EN 60068-2-27)
25 g (duration 6 ms)
2 g (duration 6 ms)
Ingress protection rating
IPx4 rated (protection against splashing water)