![Perkin-Elmer STA 6000 Installation And Hardware Manual Download Page 39](http://html1.mh-extra.com/html/perkin-elmer/sta-6000/sta-6000_installation-and-hardware-manual_1530151039.webp)
Install an STA 6000 . 39
To change the voltage:
WARNING
AVERTISSEMENT
Ensure that the mains pow er lead is disconnected from the
instrument before proceeding.
Assurez-vous que le cordon d'alimentation secteur est
débranché de l'instrument avant de continuer.
1.
Remove the fuse holder with a screwdriver by gently prying it out of the compartment.
2.
Once the holder is out far enough to hold onto, slide the fuse holder out of the slot.
3.
Remove the fuse (3.15 AT) from the 230 V position. This can be done using the point of
a pencil to gently lift the fuse out of the holder.
NOTE: It may be necessary to move the fuse slightly toward the back of the fuse pack
before prying it out of the holder.
Summary of Contents for STA 6000
Page 1: ...STA 6000 Installation and Hardware Guide THERMAL ANALYSIS ...
Page 5: ...Introduction ...
Page 7: ...Safety and Regulatory Information ...
Page 17: ...Safety and Regulatory Information 17 ...
Page 18: ......
Page 19: ...Prepare the Laboratory ...
Page 22: ...22 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...
Page 23: ...Install Multiple Analyzers ...
Page 26: ...26 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...
Page 27: ...Install an STA 6000 ...
Page 54: ...54 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...
Page 55: ...STA 6000 Hardware ...
Page 62: ...62 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...
Page 63: ...AS 8000 Autosampler ...
Page 84: ...84 STA 6000 Installation and Hardware GuideSTA 6000 Installation and Hardware Guide ...