ELECTROTHERAPY
(Applies to Performa
™
Stimul8, Quattro & Duo.)
CAUTION
• Connection of accessories other than the ones specified by the
manufacturer can adversely affect the safety of the patient and
correct functioning of the equipment, and is therefore not permitted.
• To prevent infection, electrodes should not be used on broken skin.
Before Treatment
• Check the patient for contraindications and warnings as
described on page 11.
• Test the heat sensitivity of the treatment area.
• Clean the treatment area. Any areas with excess hair should be
clipped or shaved to ensure good conductivity.
Self-Adhesive Electrodes
Self-adhesive electrodes have higher series impedance than
flexible rubber electrodes. This can cause the stimulator to
terminate treatment at higher current amplitudes. When this occurs
it is recommended to continue the treatment with flexible rubber
electrodes, properly moistened with a conductive medium.
Self-adhesive electrodes are not recommended for use with
currents that contain a DC component.
Connection and Disconnection Reactions
Constant Current (CC) output characteristics may cause
unpleasant connection and disconnection reactions if the
electrodes are not securely placed or lose contact with the skin.
Make sure the current amplitude is set to 0 mA when you apply
or remove the electrodes.