Instruction Manual peqSTAR X thermal cyclers
PEQLAB_v1212E
34
12.7.1.2
Settings
Pushing this button the adjoining dialogue
appears. The following menu items are
available: 'Programs…', 'GLP…' and
'Factory Settings'.
12.7.1.2.1
Programs Settings
Setting a check mark in this box, a 'Lid Heat'
step will be automatically inserted as first
step when generating a new program. In the
field next to 'Default LidTemp' the
temperature can be entered which shall
appear as preset value within the insertion
dialogue of a new temperature step. This
value can be changed manually in the PCR
program.
12.7.1.2.2
GLP Settings
Setting a check mark in this box, a GLP
report will be created after every run after
the query. If the box is not marked, no GLP
report will be created. No according query
will appear.
12.7.1.2.3
Factory Setting
The configurations set under Firmware >
Control Unit >Settings can be changed to the
factory settings.
12.7.1.3
Update
Updates the firmware of the control unit. The
firmware must be available on a USB
memory stick connected to the device.
Next it will be asked whether the software of control unit should be updated. If you confirm with
'Yes' the update starts.