Instruction Manual peqSTAR X thermal cyclers
PEQLAB_v1212E
9
7.7.5
Program window
In the program window the blue flashing bar is
showing the actual program step.
7.7.6
Save GLP Report
If stopping the program after the security query, the
adjoining query to create a GLP report appears.
Note: The GLP report query also appears when a
program is finished.
Will the query be confirmed by 'Yes', a GLP report
will be created and shown on the display. It will be
saved under the file name <program name>_<date>_
<time>.glp if not deleted by the user. For this the
adjoining window is shown.
The GLP report can be opened and printed in the
'GLPs' dialogue (see chapter 11).
7.7.7
Not create GLP Report
There will be no GLP report created for the aborted program. At the next run the GLP report query will
appear again. You will return to the 'Run' dialogue, which shows the last chosen program, which can be
started again.
7.7.8
Switch off the GLP report query
When you press this button, there will be no GLP reports created in the future. After the abort or end of a
program no GLP report query will appear again. The setting can be reset in the system dialogue (see
chapter 12.7.1.2).