Pulscon LTC
Function group "extended calibr." (05)
D
O
C
T
-0
0
9
4
A
0
6
/2
009
12
07
82
44
Differing from this, the following meas
u
ring error is present in the vicinity of the pro
b
e
end:
!
Note!
Please re-enter the
b
locking distance in the f
u
nction gro
u
p
"extended calibr."
(05)
f
u
nction
"upper block. dist"
(059)
when installing the device in a high nozzle:
u
pper
b
locking distance (UB) = nozzle high (H) + 50 mm.
-80
-60
-40
-20
0
20
40
60
80
0
50
100
150
200
250
300
Rod probe and coax probe
Distance from probe end [mm]
Sum of non-linearity, non-repeatability
and
hysteresis
[mm]
DC = 2
DC > 7
-80
-60
-40
-20
0
20
40
60
80
0
50
100
150
200
250
300
Distance from probe end [mm]
Sum of non-linearity, non-repeatability
and
hysteresis
[mm]
Rope probe
DC > 7
UB
H
50
After 3 s, the following message appears
E
+
–