Pulscon LTC
Function "safety settings" (01)
D
O
C
T
-0
0
9
4
A
0
6
/2
009
12
07
82
27
alarm
Instr
u
ment enters the defined alarm state (
"output on alarm"
(011)
). The alarm
message
E651
-
"level in safety distance
-
risk of overspill"
is displayed.
If the level drops o
u
t of the safety distance, the alarm warning disappears and the
instr
u
ment starts to meas
u
re again.
warning
Instr
u
ment displays a warning
E651
-
"level in safety distance
-
risk of overspill"
,
bu
t
contin
u
es to meas
u
re. If the level leaves the safety distance, the warning disappears.
self holding
Instr
u
ment switches to defined alarm state (
"output on alarm"
(011)
). The alarm
message
E651
-
"level in safety distance
-
risk of overspill"
is displayed.
If the level leaves the safety distance, the meas
u
rement contin
u
es only after a reset of
the self holding (f
u
nction:
"ackn. alarm"
(017)
).
t
[mA]
22
SD
110 %
3.6
-10 %
WARNING