Pulscon LTC
Function group "length adjustment" (03)
D
O
C
T
-0
0
9
4
A
0
6
/2
009
12
07
82
29
5
Function group "length adjustment" (03)
5.1
Function "end of probe" (030)
Use this f
u
nction to select the polarity of the pro
b
e end signal. If the pro
b
e end is
u
ncovered or in an ins
u
lated attachment, there is a negative pro
b
e end signal.
The signal from the pro
b
e end is positive if the attachment is gro
u
nded.
Selection:
•
free
•
tie down isol.
•
tie down gnd.
5.2
Function "probe length" (031)
Use this f
u
nction to select whether the pro
b
e length was changed after factory
cali
b
ration. Only then is it necessary to enter or correct the pro
b
e length.
Selection:
•
modified
•
not modified
!
Note!
If
"modified"
was selected in the
"probe length"
(031)
f
u
nction, the pro
b
e length is
defined in the next step.
5.3
Function "probe" (032)
Use this f
u
nction to select whether the pro
b
e is at the time of the commisioning
u
ncovered or covered. If the pro
b
e is
u
ncovered, the P
u
lscon LTC can determine the
pro
b
e length a
u
tomatically
"determine length"
(
034
). f
u
nction. If the pro
b
e is covered,
a correct entry is req
u
ired in the
"probe length"
(033)
f
u
nction
Selection:
•
free
•
covered