T
ROUBLESHOOTING
Blanket Warmer
▪
Operator’s Manual
▪
MN-39925
▪
Rev 2
▪
8/20
35
E-60
Real Time Clock
Checksum Error
(Blanket warmers only)
Real time clock rechargeable battery backup has discharged.
◾
Plug the warmer into the outlet for 30 minutes.
*E-61
Real Time Clock
(Blanket warmers only)
Real time clock not responding. Contact service if error persists.
E-62
Real Time Clock
(Blanket warmers only)
Timer overlay is present, but no real time clock is detected. Contact service.
*E-70
Pad Count Error
More heater pads detected than set for. Hold the standby button for 12
seconds until display shows "PAd#” (# = number of pads selected [3-7]).
Press up or down arrow to adjust to correct number of pads. [Blanket
warmers: 1.5ft
3
, 2.5ft
3
= 3 pads, 3.5ft
3
, 4.0ft
3
, 7.5ft
3
= 4 pads.
Fluid warmers: 2.5ft
3
& 4.0ft
3
= 3 pads].
*E-71
Personality Error
Contact service.
E-80
EEPROM Error
EEPROM not responding. Contact service if error persists.
*E-81
Calibration not
locked
Contact service.
*E-83
EEPROM Error
Contact service for help resetting the control.
E-87
EEPROM Error
Stored offsets corrupted. Offsets reset to 0.
◾
Control may need to be re-calibrated.
◾
Possible bad EEPROM.
◾
Contact service if error persists.
E-90
Button stuck
A button has been held down for >60 seconds. Adjust control. Error will
reset when the problem has been resolved.
E-95
Factory Test pin
short detected.
Ensure that debris is not causing a short between the test pins. If the pins
are clean, replace the control.
*E-98
Temperature
Delta Error
Temperature of the cavity sensors 1 and 2 differ by more than 3.3ºC (6ºF).
◾
Remove product and allow the warmer to cool down.
◾
Verify that the product sensor is clean and operating correctly.
◾
Set the power switch to the OFF position to clear the error code.
◾
If error persists, the cavity sensor should be replaced by a qualified
service technician. Contact service.
E-99
Hardware Over
Temp
◾
Inspect the connections and condition of high limit bimetal thermostat.
◾
If applicable, make sure that the compartment fan motor is operating.
◾
Contact service if error persists.
Code
Refers to
Action required
Summary of Contents for P-2040
Page 1: ...Operator s Manual Blanket Warmer P 2040 P 2055 MN 39925 REV 02 8 20 EN ...
Page 38: ......
Page 39: ......