background image

S

PECIFICATIONS

Blanket Warmer   

   Operator’s Manual   

   MN-39925   

   Rev 2   

   8/20

15

P-2055

  

EC-DIM-004924

 

 

Adjustable shelf

56.0" (1422mm)

29.7" (754mm)

30.2" (767mm)

23.75" (603mm)

29.25" (743mm)

Chamber

64.1" (1629mm)

71" (1801mm)

18-1/4" (465mm)

Electrical outlet

6.8"

Cord length
8' (2,438mm) (est.)

Adjustable shelf

Chamber

32" (811mm)

26.5" (672mm)

68.75" (1746mm)

24.0" (610mm)

Chamber

Electrical outlet

56.7" (1440mm) 

(172mm)

Summary of Contents for P-2040

Page 1: ...Operator s Manual Blanket Warmer P 2040 P 2055 MN 39925 REV 02 8 20 EN ...

Page 2: ...eproduced or used in any manner whatsoever without the express written permission of Pedigo Products Inc Trademarks All trademarks referenced in this documentation are the property of their respective owners Manufacturer Pedigo Products Inc 4000 S E Columbia Way Vancouver WA 98661 USA 1 800 822 3501 www pedigo usa com Original instructions The content in this manual is written in American English ...

Page 3: ... most durable convenient efficient and safe warming equipment on the market All warmers are manufactured and fully inspected in the USA with a commitment to quality Pedigo Repair Service Call Call 262 509 6517 to reach our service call center for immediate access to local authorized service agencies This service is provided exclusively for Pedigo Product Inc equipment and is available throughout t...

Page 4: ...FOREWORD Blanket Warmer Operator s Manual MN 39925 Rev 2 8 20 4 This page intentionally left blank ...

Page 5: ...Drawings 14 Installation 17 How to Unpack the Warmer 17 How to Install the Warmer 18 Operation 19 Preparing the Warmer for First Use 19 How to Turn On and Turn Off the Warmer 20 How to Operate the Blanket Warmer 21 How to Use the Timer 22 How to Change the Temperature Scale 24 How to Change the Sound Settings 25 How to Lock and Unlock the Controller 26 Maintenance 27 Maintenance Schedule 27 How to...

Page 6: ...TABLE OF CONTENTS Blanket Warmer Operator s Manual MN 39925 Rev 2 8 20 6 This page intentionally left blank ...

Page 7: ...ANGER Danger indicates a hazardous situation which if not avoided will result in serious injury or death WARNING Warning indicates a hazardous situation which if not avoided could result in serious injury or death CAUTION Caution indicates a hazardous situation which if not avoided could result in minor or moderate injury NOTICE Notice indicates a situation which if not avoided could result in pro...

Page 8: ... Do not use this warmer for warming blood or blood products Do not cover or block any of the openings of this warmer Do not use this warmer in a wet location Only clean the warmer when the power cord is unplugged Do not use corrosive chemicals when cleaning the warmer Do not use the warmer cavity for storage Do not remove exterior panels from the warmer or attempt repairs The warmer has no user se...

Page 9: ...ed to or shipped with the warmer Only trained personnel are permitted to service or repair the warmer Repairs that are not performed by a trained technician or the use of non factory parts will void the warranty and relieve all liability Any troubleshooting guides and components views included with this manual are for reference only and are intended for use by qualified and trained service technic...

Page 10: ...SAFETY Blanket Warmer Operator s Manual MN 39925 Rev 2 8 20 10 This page intentionally left blank ...

Page 11: ...LABELS Blanket Warmer Operator s Manual MN 39925 Rev 2 8 20 11 L ABELS Label Locations EC PHD 004935 a c d b ...

Page 12: ...y seal 2 Electrical warning label 3 Pedigo contact information 4 Quality inspected S e c u r i t y S e a l S e c u r i t y S e a l S e c u r i t y S e a l S e c u r i t y S e a l S e c u r i t y S e a l S e c u r i t y S e a l S e c u r i t y S e a l S e c u r i t y S e a l ...

Page 13: ...d relative humidity is above 20 non condensing Standards for electrical equipment Medical equipment listed by Underwriters Laboratories with respect to electrical shock fire and mechanical hazards only in accordance with UL 60601 1 and CAN CSA C22 2 No 601 1 Grounding reliability can only be achieved when the appliance is connected to an equivalent receptacle marked Hospital Grade IP XO ordinary M...

Page 14: ...10mm 26 8 681mm Compartment 26 8 681mm Compartment 66 4 1686mm 71 1803mm 54 0 1372mm 10 9 276mm 6 8 172mm 32 5 826mm 28 8 731mm 61 4 1560mm 13 5 342mm Insert 9 8 248mm Insert 18 8 478mm Compartment 16 9 430mm Insert Cord length 8 2 438mm est 21 6 548mm Radius 13 5 342mm Insert 9 8 248mm Insert 26 6 676mm Compartment ...

Page 15: ... 004924 Adjustable shelf 56 0 1422mm 29 7 754mm 30 2 767mm 23 75 603mm 29 25 743mm Chamber 64 1 1629mm 71 1801mm 18 1 4 465mm Electrical outlet 6 8 Cord length 8 2 438mm est Adjustable shelf Chamber 32 811mm 26 5 672mm 68 75 1746mm 24 0 610mm Chamber Electrical outlet 56 7 1440mm 172mm ...

Page 16: ...SPECIFICATIONS Blanket Warmer Operator s Manual MN 39925 Rev 2 8 20 16 This page intentionally left blank ...

Page 17: ...e box or crate Save all packing materials for inspection by the carrier 2 Cut and remove the retaining straps and plastic wrap 3 Remove the warmer from the pallet using an appropriate lifting device 4 Remove the paperwork from the cavity Result The warmer is now unpacked WARNING Crushing hazard An unstable warmer can lead to a crushing hazard Observe your facility s best practices for moving large...

Page 18: ... the following Step Action 1 Make sure that The location where the warmer is being installed is rated to support the weight of the warmer The warmer is within five feet of the appropriate electrical outlet You follow the warmer clearance guidelines Refer to topic Specification Information for clearance requirements 2 Move the warmer to the installation location and onto the final resting surface 3...

Page 19: ...ved cleaning agent if painted panels 3 Clean the interior of the warmer with a damp cloth or approved cleaning agent 4 Dry the interior and exterior of the warmer with a clean lint free cloth Leave the door open until the interior of the warmer completely dries 5 Clean each side of the window pane with glass cleaner or distilled vinegar Result The warmer is now ready for use CAUTION Burn hazard Al...

Page 20: ... OFF button The screen turns on The warmer is now on Turning off the warmer To turn off the warmer do the following 3 Press and hold the ON OFF button until the screen turns off then release the button The warmer is now off NOTE A temporary odor may be noticeable upon initial start up of the warmer Contact the manufacturer if the odor persists after a day or more of continuous use EC TS 004951 a N...

Page 21: ...Rotate the blankets daily from the bottom to the top of the load to ensure equal heating and to prevent discoloration Result The blankets are now warming WARNING Personal injury hazard Do not operate the warmer in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide in oxygen enriched environments or in any other potentially explosive environment WARNING Personal injury ...

Page 22: ... the TIMER button The TIME icon a illuminates 2 Press the up and down arrow buttons to change the time of day 3 Press the TIMER button again The START icon b illuminates Press the up and down arrow buttons to change the start time 4 Press the TIMER button again The STOP icon c illuminates Press the up and down arrow buttons to change the stop time 5 Press the TIMER button again to enable the timer...

Page 23: ...mer Operator s Manual MN 39925 Rev 2 8 20 23 Continued from previous page Disabling the timer 6 Press and hold the TIMER button until the timer status LED indicator goes out Result The timer has now been enabled or disabled ...

Page 24: ... F to C and vice versa do the following Step Action 1 Press and hold the ON OFF button until the screen turns off then release the button 2 Press and hold the up arrow button until the warmer beeps and displays the temperature scale then release the button 3 Press the up and down arrow buttons to toggle between the temperature scales 4 Press the ON OFF button to set the temperature scale Result Th...

Page 25: ...ress and hold the ON OFF button until the screen turns off then release the button 2 Press and hold the down arrow button until the warmer beeps and displays the sound setting of 0 or 1 then release the button 3 Press the up and down arrow buttons to turn the sound on or off 4 Press the ON OFF button to set the sound setting Result The sound setting has now been changed NOTE Setting the sound to 0...

Page 26: ...e following Step Action 1 Press and while holding the ON OFF button press the up arrow button The lock indicator will illuminate Unlocking the controller To unlock the controller do the following 2 Press and while holding the ON OFF button press the down arrow button The lock indicator goes off Result The controller has now been locked or unlocked NOTE Make sure the set point temperature did not c...

Page 27: ...ance do the following Check the door gasket for tears and holes Make sure that it is firmly attached to the door Check the seal when the door is closed the guards around the air temperature sensor are in place and fully secure to the warmer the hardware securing the warmer s to the wall if applicable Clean the outside of the warmer with stainless steel cleaner See topic How to Clean the Warmer for...

Page 28: ...e with a thermocouple placed 1 25mm from the cavity sensor Do not allow the thermocouple to touch any surface Monitor the temperature for approximately one hour in an empty cavity the condition of the plug and cord and replace if damaged the controller screen for excessive wear Make sure the controller screen displays and operates properly Clean the shelves and interior of the warmer See topic How...

Page 29: ...le 4 Clean the interior of the warmer with a damp cloth 5 Dry the interior and exterior of the warmer with a clean lint free cloth Leave the door open until the interior of the warmer completely dries 6 Clean each side of the window pane with glass cleaner or distilled vinegar Continued on next page WARNING Electric shock hazard Disconnect the warmer from electric power before cleaning CAUTION Bur...

Page 30: ...the support assembly and interior with a damp cloth 4 Dry the interior of the warmer with a clean lint free cloth Leave the door open until the interior of the warmer completely dries 5 Clean each side of the window pane with glass cleaner or distilled vinegar 6 Re install the support assembly Result The warmer is now clean WARNING Personal injury hazard The blankets may overheat if the shelves an...

Page 31: ...unds an alarm and the screen displays an error code when the warmer malfunctions Procedure If the warmer malfunctions do the following Step Action 1 Press the ON OFF button to acknowledge the alarm Refer to the troubleshooting section for the error code 2 Make sure the blankets are the correct temperature before using Result Resume operation of the warmer ...

Page 32: ... can be manually reset using the high limit manual reset button s located at the rear of the warmer Procedure To manually reset the warmer do the following Step Action 1 Press the high limit manual reset button s a Result Resume operation of the warmer NOTE If the manual reset warmer button s trips again while the warmer is operating contact Pedigo for the authorized service partner in your area E...

Page 33: ...rmer to the off O position and allowing the warmer to cool NOTICE Do not attempt to repair or service the warmer beyond this point Contact the manufacturer for the nearest authorized service agent Repairs made by any other service agent without prior authorization by the manufacturer will void the warranty Code Refers to Action required Display flashes set point Cavity temperature higher than set ...

Page 34: ...m the warmer control PCB and use an Ohm meter to measure the resistance of the sensor Sensor PR 37140 should measure 10 KOhm 1 5 KOhm at 25 C 77 F Sensor SN 33541 should measure 100 Ohm 10 Ohm at 25 C 77 F Replace sensor if measurement is outside of tolerance E 31 Cavity sensor The sensor reading is above the temperature set point Blanket warmer triggers at 15 over set point Fluid warmer triggers ...

Page 35: ...ntact service E 83 EEPROM Error Contact service for help resetting the control E 87 EEPROM Error Stored offsets corrupted Offsets reset to 0 Control may need to be re calibrated Possible bad EEPROM Contact service if error persists E 90 Button stuck A button has been held down for 60 seconds Adjust control Error will reset when the problem has been resolved E 95 Factory Test pin short detected Ens...

Page 36: ... Check to make sure the fan sensor wires did not disconnect from the fan or the control board If the error persists after checking the wires replace the fan Flashing yellow iv fluid temperature status indicator Flashing yellow LED is illuminated within iv storage shelf Set the power switch to the OFF position Remove any iv bags from inside the warmer Set the power switch to the ON position Visuall...

Page 37: ...ontaining chloride or quaternary salts poor water quality or equipment with missing or altered serial numbers Any losses or damage resulting from malfunction including loss of contents or consequential or incidental damages of any kind Equipment modified in any manner from original model substitution of parts other than factory authorized parts removal of any parts including legs or addition of an...

Page 38: ......

Page 39: ......

Page 40: ...digo Products Inc 4000 S E Columbia Way Vancouver WA 98661 Phone 800 822 3501 Fax 360 696 1700 www pedigo usa com Made in the U S A Printed in the U S A Specifications are subject to change without notice ...

Reviews: