Fig. 103
Cursor time measurement
Cursor measure position menu as below:
Function
Display
Instruction
M1-M2
Position
The position difference between two cursors in memory area.
Cursor 1
Position
The position of cursor 1 corresponds to trigger in memory area.
Cursor 2
Position
The position of cursor 2 corresponds to trigger in memory area.
Follow up below steps to measure data width of pulse with cursor measurement
1.
Press “
9(CURSOR)
” and Cursor menu appears.
2.
Press “
F1
” till cursor display as “
Position
”. Two purples vertical lines display in panel and refer to Cursor
1 and Cursor 2 separately.
3.
Turn “
CH1 POSITION
” knob and set Cursor 1 position in falling edge of CH00
4.
Turn “
CH2 POSITION
” knob and set Cursor 2 position in next falling edge of CH00.
5.
Now the position difference between two cursors display in “
F2
” is “
-10
” which means there is 10 acquired
data between two cursors. The position of Cursor 1 corresponds to trigger in memory position display in
“F3” as “
2097158
”. The position of Cursor 2 corresponds to trigger in memory position display in “F4” as
“
2057168
”.
Fig. 104
Cursor position measurement