34. How to Save and Recall a Wave Form
Press the
SAVE/RCL
button, you can save and call out the waveforms in the instrument. The menu displayed
in the screen is shown as Fig.59.
Fig. 59
Wave Form Save Menu
The description of the
Wave Form Save Menu
is shown as the following table:
Function Menu
Setting
Description
Source
CH1
CH2
MATH
Choose the wave form to be saved.
WAVE
A , B
C , D
Choose the address in or from which the waveform
is saved or can be get access to.
Save
Save the wave form of the source in the selected
address.
CH (X)
X=A,B,C oder D
OFF
ON
Switch on or off the display of the stored waveform.
The
PeakTech
®
1190 / 1230 oscilloscopes can store four waveforms, which can be displayed with the current
waveform at the same time. The stored waveform called out cannot be adjusted.
In order to save the waveform of the channel CH1 into the address A, the operation steps should be followed:
1.
Press the
F1
menu selection button and choose CH1 for Source.
2.
Press the
F2
menu selection button and choose A as save address.
3.
Press the
F3
menu selection button and save the waveform.
Press the
F4
menu selection button and choose
ON
for CH A. The stored wave form A will be displayed in the
screen.
-173-