Alternate trigger (Trigger Mode:video)
Alternate trigger (Trigger Type:video) Menu is shown as Fig. 45.
Fig. 45
Alternate trigger (Trigger Type:video) Menu
Alternate trigger (Trigger Type: video) Menu list:
MENU
SETTING
INSTRUCTION
Source
CH1
CH2
EXT
EXT/5
Select CH1 as the trigger source.
Select CH2 as the trigger source.
Ext-trigger
Ext-trigger divide to 5 to extend trigger level range
Mode
Video
Sync
Line
Field
Odd Field
EvenField
Designed Line
Synchronic trigger in video line.
Synchronic trigger in video field.
Synchronic trigger in video odd filed
Synchronic trigger in video even field
Synchronic trigger in designed video line
Modulation
NTSC
PAL/SECAM
Video modulation setting
Holdoff
100ns
~
10s
Turn “TRIG LEVEL” knob to set time slot before another
trigger event
Holdoff
Reset
Reset hold time to 100ns
-163-