28.5. Alternate trigger
Trigger signal comes from two vertical channels when alternate trigger is on. This mode is used to observe two
unrelated signals. You can choose different trigger modes for different channels. The options are as follows:
edge, video, pulse or slope.
Alternate trigger (Trigger mode: Edge)
Alternate trigger (Trigger Type: Edge) Menu is shown as Fig.44.
Fig. 44
Alternate trigger(Trigger Type: Edge) Menu
Alternate trigger(Trigger Type: Edge) Menu list:
MENU
SETTING
INSTRUCTION
Source
CH1
CH2
Select CH1 as the trigger source.
Select CH2 as the trigger source.
Mode
Edge
Set vertical channel trigger type for edge trigger.
Slope
Rising
Falling
Trigger in signal rising edge
Trigger in signal falling edge
Coupling
AC
DC
HF
LF
Not allow DC portion to pass.
Allow all portion pass.
Not allow high frequency of signal pass and only low frequency portion
pass.
Not allow low frequency of signal pass and only high frequency portion pass
Holdoff
100ns
~
10s
Turn “TRIG LEVEL” knob to set time slot before another trigger event.
Holdoff
Reset
Reset hold time to 100ns
-161-