
29. How to Operate the Function Menu
The function menu control zone includes 7 function menu buttons and 3 immediate-execution buttons:
SAVE/RCL, MEASURE, ACQUIRE, UTILITY, CURSOR, DISPLAY, AUTOSCALE, AUTOSET, RUN/STOP
and U-DISK COPY.
30. How to Implement Sampling Setup
Press the
ACQUIRE
button and the menu is displayed in the screen, shown as Fig. 49.
Fig. 49
ACQU MODE Menu
The description of the
Sampling Setup Menu
is shown as follows:
Function Menu
Setting
Description
Sample
General sampling mode.
Peak Detect
It is used for the detection of the jamming burr and the
possibility of reducing the confusion.
Average
It is used to reduce the random and don’t-care noises, with
the optional number of averages.
Averages
4, 16, 64, 128
Choose the number of averages.
Change the
ACQU Mode
settings and observe the consequent variation of the wave form displayed.
-167-