6.7. Battery Test
1.
Insert the black test lead banana plug into the negative COM jack
and the red test lead banana plug into the positive V jack.
2.
Select the 1,5V or 9V BAT position using the function select
switch.
3.
Connect the red test lead to the positive side of the 1,5V or 9V
battery and the black test lead to the negative side of the 1,5V or
9V battery.
4.
Read the voltage in the display.
Good
Weak
Bad
9 V battery
>8,2 V
7,2 to 8,2 V
<7,2 V
1,5 V batttery
>1,35 V
1,22 to 1,35 V
<1,22 V
6.8. Non-Contact Voltage (NCV)
The NCV function works on any rotary switch position.
1.
Test the detector on a known live circuit before use.
2.
Hold the top of the meter very close to the voltage source as
shown.
3.
If voltage is present, the rim of the LCD display will flash a bright
red.
Note:
The voltage detector operates from a voltage > 50V / 50Hz.
-35-