Note:
1.
If the resistance value being measured exceeds the maximum
voltage of the range selected, an over-range indication will be
displayed "OL". Select a higher range. For resistance approx. 1
MΩ and above, the meter may take a few seconds to stabilize.
This is normal for high resistance readings.
2.
When the input is not connected, i. e. at open circuit, the
figure "OL" will be displayed for the overrange condition.
3.
When checking in-circuit resistance, be sure the circuit under test
has all power removed and that all capacitors are fully discharged.
6.6. Diode Measurement
1.
Connect the black test lead to the COM jack and the red test lead
to
the V/Ω/-mA-jack. (Note: the polarity of the red test lad is "+")
2.
Set the function switch to the
-range.
3. Connect the test leads across the diode under measurement.
Note:
1.
When the input is not connected, i. e. at open circuit, the
figure "OL" will be displayed for the overrange condition.
2.
There is 1 mA current flow through the device under test
3.
The meter displays the forward voltage drop in millivolts,
and overload when the diode is reversed.
-34-