* The meter is designed to withstand the stated max voltages. If it is not
possible to exclude without that impulses, transients, disturbance or
for other reasons, these voltages are exceeded a suitable presale
(10:1) must be used.
* Replace a defective fuse only with a fuse of the original rating. Never
short-circuit fuse or fuse holding.
* Disconnect test leads or probe from the measuring circuit before
switching modes or functions.
* Do not conduct voltage measurements with the test leads connected
to the mA/A- and COM-terminal of the equipment.
* The 10A-range is protected. To avoid damage or injury, use the meter
only in circuits limited by fuse or circuit breaker to 10A or 2000VA.
* To avoid electric shock, disconnect power to the unit under test and
discharge all capacitors before taking any resistance measurements.
* Do not conduct current measurements with the leads connected to the
V/
-terminals of the equipment.
* Check test leads and probes for faulty insulation or bare wires before
connection to the equipment.
* Please use only 4mm-safety test leads to ensure immaculate function.
* To avoid electric shock, do not operate this product in wet or damp
conditions. Conduct measuring works only in dry clothing and rubber
shoes, i. e. on isolating mats.
* Never touch the tips of the test leads or probe.
* Comply with the warning labels and other info on the equipment.
* The measurement instrument is not to be to operated unattended.
* Always start with the highest measuring range when measuring
unknown values.
* Do not input values over the maximum range of each measurement to
avoid damages of the meter.
* Do not subject the equipment to shocks or strong vibrations.
* Do not subject the equipment to direct sunlight or extreme
temperatures, humidity or dampness.
* Keep hot soldering irons or guns away from the equipment.
* Allow the equipment to stabilize at room temperature before
taking up measurement (important for exact measurements).
-22-