![PeakTech 1015 GPIB Operation Manual Download Page 36](http://html1.mh-extra.com/html/peaktech/1015-gpib/1015-gpib_operation-manual_1517473036.webp)
31
FRQ
Command Execution examples
Example 1. To set 1kHz
You would send the following program message:
FRQ1kHz
Example2. To set 10kHz
You would send the following program message:
FRQ10kHz
Note:
* When you enter FRQ1.5kHz, the instrument generates 1.500kHz automatically.
* When you enter FRQ1.5MHz, the instrument generates 1.500MHz automatically.
More in one line
It is allowed to combine multiple commands and/or queries with their respective data into one single line.
To combine multiple commands and queries with their respective data, a separator must be used. The
semicolon (
;
) is used as a separator between commands and/or queries. This separator is officially
called a “program message unit separator”. The followings are the examples to combine multiple
commands and/or queries.
Example1. To generate
sine
waveform,
1kHz
You would send the following program message:
WFM0;FRQ1kHz
Example2. To generate
sine
waveform,
1kHz
,
Amp
litude
50
You would send the following program message:
WFM0;FRQ1kHz;AMP50
All rights, also for translation, reprinting and copy of this manual or parts are reserved.
Reproduction of all kinds (photocopy, microfilm or other) only by written permission of the publisher. This
manual considers the latest technical knowing. Technical changing which are in the interest of progress
reserved.
We herewith confirm, that the units are calibrated by the factory according to the specifications as per
the technical specifications.
We recommend to calibrate the unit again, after 1 year.
PeakTech
®
05/2004