Start-Up Sequence
Switch the generator on using the power switch on the rear panel of the unit. Please
note: Air will only flow through the generator when the power is switched on. The LED’s
on the front panel will illuminate
YELLOW
for the start-up period.
The generator will purge impurities from its internal storage tank until Oxygen content
is less than 5PPM, approximately 1.5 hours (3-3.5 hours for Headspace) of uninterrupted
operation. On completion, the Fascia will illuminate
GREEN
or
BLUE (70-**** Part
Numbers Only)
and the generator is now ready to be connected to the application.
On Demand Gas
The generator will produce nitrogen on demand. If the application is operating
and requires a gas flow, the system it is connected to will supply this to suit the
requirements of the application. If the application requirement for gas stops, the system
will also stop, once it has reached its upper set limit in the internal storage tanks. If the
demand from the application starts again, the system will detect the demand for gas
and will automatically start again to suit the demand.
Unusual Operation
If at any time the generator begins to emit excessive noise or vibration, then it should
be switched off and you should contact Peak Scientific or the Peak Partner from which
the generator has been purchased.
YELLOW for 90 Minutes
GREEN / BLUE