Page 10
Technical Specification
Precision Nitrogen Headspace
Environment
Minimum Operating Ambient Temperature
5°C (41°F)
Maximum Operating Ambient Temperature
35°C (95°F)
Maximum Altitude
2000 m
Maximum Relative Humidity
70% Non-Condensing
Minimum Storage Temperature*
-20°C (-4°F)
Maximum Storage Temperature*
60°C (140°F)
Inlet Conditions
Inlet Air Pressure Min-Max
8.3-8.6 bar (120-125 psi)
Minimum Air Inlet Flow
18 l/min
Minimum Air Inlet Quality
ISO 8573-1:2010 Class[1.4.1]
Particles
<0.01µm
Generator Outlets
Maximum Gas Output Pressure
6.9 bar (100 psi)
Maximum Outlet Flow Rate
250cc
Start-up time
3 - 3.5 Hours
Particles
<0.01µm
Phthalates
NONE
Suspended Liquids
NONE
Gas Outlets
1 x ¼” BSPP
For optimum performance from your Precision Nitrogen generator, the inlet gas supply pressure should be 30PSI greater than the
required output pressure.
Though the Precision Headspace Generator delivers nitrogen at a purity suitable for most non- trace applications, we would
recommend the installation of oxygen and hydrocarbon traps for analysis at lower detection limits (<1000ppm).
Electrical Requirements
Voltage
110-230v ±10%
Frequency
50/60 Hz
Current
0.37 -0.17 A
Input Connection
C14 Plug
Power Cord (Supplied)
C13 Socket to Local Connection
Pollution Degree
2
Insulation Category
II
General
Dimensions cm (inches) WxDxH
38.0 (14.9) x 54.0 (21.2) x 25.6 (10.0)
Generator Weight Kg (lbs)
21 (46.2)
Shipping Weight Kg (lbs)
24 (53)