Page 11
Technical Specification
Precision Nitrogen Trace
Environment
250cc
600cc
1000cc
Minimum Operating Ambient Temperature
5°C (41°F)
Maximum Operating Ambient Temperature
35°C (95°F)
Maximum Altitude
2000 m
Maximum Relative Humidity
70% Non-Condensing
Minimum Storage Temperature*
-20°C (-4°F)
Maximum Storage Temperature*
60°C (140°F)
Inlet Conditions
Inlet Air Pressure Min-Max
8.3-10 bar (120-145 psi)
Minimum Air Inlet Flow
18 l/min
22 l/min
Minimum Air Inlet Quality
ISO 8573-1:2010 Class[1.4.1]
Particles
<0.01µm
Generator Outlets
Maximum Gas Output Pressure
5.5 bar (80 psi)
Maximum Outlet Flow Rate
250cc
600cc
1000cc
Start-up time
1.5 Hours
Particles
<0.01µm
Phthalates
NONE
Suspended Liquids
NONE
Gas Outlets
1 x ¼” BSPP
Electrical Requirements
Voltage
115 VAC ±10%
230VAC ±10%
Frequency
60 Hz
50 Hz
Current
4.4A
2.1A
Input Connection
C14 Plug
Power Cord (Supplied)
C13 Socket to Local Connection
Pollution Degree
2
Insulation Category
II
General
Dimensions cm (inches) WxDxH
38.0 (14.9) x 54.0 (21.2) x 40.6 (15.8)
Generator Weight Kg (lbs)
38 (83.6)
38 (83.6)
36(79.2)
Shipping Weight Kg (lbs)
42 (92.4)
42 (92.4)
40(88)