background image

Genius XE 

(All Models) 

User Manual

Copyright © Peak Scientific Instruments Ltd 2020 — GeniusXE User Manual  Rev 9  RSID 2299  EN —18/02/2020

Summary of Contents for Genius XE 70

Page 1: ...Genius XE All Models User Manual Copyright Peak Scientific Instruments Ltd 2020 GeniusXE User Manual Rev 9 RSID 2299 EN 18 02 2020...

Page 2: ...ranty your generator must be registered to the end user not a reseller or distributor and must have a paid annual preventative maintenance arranged within 12 months of the installation date carried ou...

Page 3: ...13 Generator Overview 14 Drain Connection 16 Electrical Connection 17 Start Up Sequence 18 Product Registration 19 Connecting to the application 21 Normal Operation 23 Eco Mode 23 Service Screens 24...

Page 4: ...Update Liam Couttie 16 12 2019 9 Technical Specification Update Liam Couttie 18 02 2020 How to use this Manual This manual is intended for end users and has been written as a reference document where...

Page 5: ...upply to the generator must comply with ISO 8573 1 2010 Class 1 2 1 at all times d your air compressor dryer filtration and oil removal systems must be deemed suitable for use by Peak or a Peak Partne...

Page 6: ...conditions outlined in this warranty state ment in any respect or to create any additional obligations or liabilities for any party involved 11 This warranty statement supersedes any and all prior war...

Page 7: ...ted conditions are fully understood or met A CAUTION notice denotes a hazard It calls attention to an operating procedure process or similar which if not correctly performed or adhered to could cause...

Page 8: ...nt for measurement control and laboratory use Electromagnetic Compatibility Directive 2014 30 EU EN 61326 1 2013 Electrical Equipment for measurement control and laboratory use FCC47 CFR Part 15 Class...

Page 9: ...h confirm that as a manufacturer of articles we conform to the European Parliament and Council Directives 2011 65 EU for the original 6 restricted substances listed in the Restriction of Hazardous Sub...

Page 10: ...0V 10 230V 10 Frequency 120V 60Hz 230V 50 60Hz 230V 50 60Hz Current 12A 120V 8A 230V 12A 230V Input Connection C20 Plug Power Cord C19 Socket Circuit Breakers 230V 6A MCB 2A MCB 120V 10A MCB 2A MCB Po...

Page 11: ...the Peak Partner from where the unit was purchased Follow the unpacking instruction supplied with the generator It will require two people to remove the unit from the shipping crate and to manoeuvre...

Page 12: ...1 8 US 120V Mains Power Cable x 1 9 Genius XE Installation Guide x 1 Genius XE 35 XE 70 230V 1 1 4 x 1 4 Compression Fitting x 1 2 1 4 x 6mm Push Fit Fitting x 2 3 Flow Control Silencer x 1 4 1 4 PTF...

Page 13: ...et could adversely affect its performance Consideration should also be given to the air flow around the unit An air gap of 75mm 3 should be maintained down both sides and 100mm 4 at the rear of the un...

Page 14: ...mensions Genius XE 35 Genius XE 70 The generator must always be placed on a flat level surface Failure to do so will affect the performance of the generator WARNING 570 mm 22 4 710 mm 27 9 650 mm 25 6...

Page 15: ...ear Connections Ensure outlets are connected to correct applications Connections should only be carried out by trained personnel Unit Controls Touch Screen Nitrogen Port Drain Port Power Socket WARNIN...

Page 16: ...ain connection or container It should be noted that the generator can expel a considerable amount of water dependant on ambient humidity If a container is used it should be emptied at regular interval...

Page 17: ...te The mains supply must be of the stated AC voltage and frequency EARTH GROUND E Green Yellow or Green LIVE L Brown or Black Neutral N Blue or White Electrical requirements are 120VAC nominal 5 or 23...

Page 18: ...and switching it on for the purge run it is necessary to fit the silencer to the nitrogen output port Once this is done the generator can be connected to the mains and switched on The Genius XE will t...

Page 19: ...also be requested by phoning the Peak helpdesk This code is only required on the initial start up of the unit 1 Press the I ve got my pin button on the screen on the front of the generator 2 Enter you...

Page 20: ...Page 20 4 Upon completion of the initialisation phase the screen will display the below home screen GENIUS XE 35 S N XXXXXXXXXXXX Nitrogen XXXX psi...

Page 21: ...lease ensure that it is thoroughly checked for being leak tight Even the slightest leak in the gas supply between the generator and the application can lead to a drop in Nitrogen purity or insufficien...

Page 22: ...116psi The output pressure can be adjusted using the pressure regulator at the rear of the generator and viewed on the adjacent gauge To ensure a smooth pressure profile it is not recommended to incre...

Page 23: ...5 nitrogen at the maximum rated output flow Eco Mode To reduce energy consumption and minimise component wear Genius XE generators feature a smart Eco Mode which ensures gas is produced as is necessar...

Page 24: ...XXXXXXXXXXX Nitrogen XXXX psi Service Due Home Screen Warning This is the main home screen when a service is due displaying a yellow service warning The screen will alternate between the home screen a...

Page 25: ...ng as shown next GENIUS XE 35 S N XXXXXXXXXXXX Nitrogen XXXX psi REF CODE XXXX REF CODE XXXXXXXXXX OR VISIT WWW PEAKSCIENTIFIC COM SERIAL XXXXXXXXXXXX Confirm Minor Error Warning Screen This is the fu...

Page 26: ...100 l min Temp Hum Temp Hum 10 xx xx xx xx xx xx xx xx xx xx xx xx xx xx 0 10 20 30 40 C 50 60 70 80 90 0 10 20 30 40 50 60 70 80 90 100 Temperature Graph 2 results plotted Temperature Humidity Y axi...

Page 27: ...ct Pressure Unit bar atm kPa Select Flow Unit l min l hour m3 min m3 hour Select Flow Unit Screen Allows the user to select a unit of Flow Select Pressure Unit Screen Allows the user to select a unit...

Page 28: ...l Component Visit Qty 12 Months Genius XE 35 120 230V Annual Maintenance Kit www peakscientific com ordering 1 Genius XE 70 230V Annual Maintenance Kit 1 48 Months Genius XE 35 120V 48 Month Kit 1 Gen...

Page 29: ...essing the icon will show the message again Stage 2 If the service is not completed the generator will continue to run After 2 weeks the service overdue indicator will show on the HMI along with a war...

Page 30: ...to install and setup your generator IQ OQ Certified assurance for applications requiring documented qualification Complete plan Swift response by a Peak Service Engineer within 72 hours planned preve...

Page 31: ...he cloth prior to use Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the generator Under no circumstances should any solvents or abrasive cl...

Page 32: ...d contact your service provider Minor Alarm Message There is an issue with the generator but it will continue to supply gas at the required pressure flow and purity Major Alarm Message There is a prob...

Page 33: ...tact your service provider Service indicator on the screen is active The generator is due for service Contact your service provider Refer to Service Indication section of this manual for further infor...