6 Camera Control for the pco.dimax
pco.dimax User’s Manual V1.00 © PCO AG, Germany
15
6.7
The “I/O Signals” Sheet
For corresponding interfaces please see rear site of camera:
Acquire Enable
If checked, a signal for ‘Acquire Mode’ (see chapter 6.6) is accepted.
Set Ready
Feature not yet implemented.
Exposure Trigger
If checked, a signal for [External Exp. Start] and [External Exp. Ctrl] (see
chapter 6.2) is accepted.
Status Expos / Status Busy
If checked, a signal indicating exposure status or busy status is given at
respective BNC output.
[Type]
TTL:
Maximum low level: 0.8V
Minimum high level: 2V
High Lev.:
Maximum low level: 5V
Minimum high level: 10V
CONTACT:
Maximum switch on resistance: 100Ohms
Minimum switch off resistance: 40kOhms
Minimum switch voltage rating: 5VDC
Minimum switch current rating: 1mA
RS485:
Receiver sensitivity: +-200mV
Note: termination of 120Ohms outside camera required
[Filter]
Feature not yet implemented.
[Polarity]
active for high/low signal or rising/falling edge