© PCE Instruments
42
6
Technical Specifications
Voltage DC V
Measuring range
Resolution
Accuracy
0 ... 30V
0.001V
± 0.02% v.Mw. + 2 Dgt
Current DC mA
Measuring range
Resolution
Accuracy
0 ... 24-mA
0.001-mA
± 0.02% v.Mw. + 2 Dgt
Measurement / Simulation
Thermocouples
Measuring range
Resolution
Accuracy
e
-200 ... 1200°C / -328 ...
2192°F
0.1°C
/
0.18°F
± 0.3°C / 0.54°F
J
-200 ... 1200°C / -328 ...
2192°F
0.1°C /
0.18°F
± 0.3°C / 0.54°F
K
-200 ... 1372°C / -328 ...
2501°F
0.1°C /
0.18°F
± 0.3°C / 0.54°F
T
-200 ... 400°C / -328 ...
752°F
0.1°C /
0.18°F
± 0.3°C / 0.54°F
B
450 ... 1800°C / 842 ...
3272°F
0.1°C /
0.18°F
± 0.3°C / 0.54°F
R
0 ... 1750°C / 32 ...
3182°F
0.1°C /
0.18°F
± 0.3°C / 0.54°F
S
0 ... 1750°C / 32 ...
3182°F
0.1°C /
0.18°F
± 0.3°C / 0.54°F
N
-200 ... 1300°C / -328 ...
2372°F
0.1°C /
0.18°F
± 0.3°C / 0.54°F
mV
10 ... 80 mV
-10 ... 250 mV
0.001 mV
0.01 mV
± 0.02% of the displayed
value + 2μV
± 0.02% of reading + 0.02
mV
Note:
The specifications refer to the temperature standard ITS-90