© PCE Instruments
36
File No.
File Number.
Range: 1 … 25
This parameter is enabled only for Periodic Trigger.
File Info.
Shows the information of stored files. This information contains Logging
Start Time & Date and No of Samples stored in the file.
File Delete
Delete stored file.
START
LOGGING
Press F1 key while selecting this button to start the Logging.
Notes
:
•
Maximum No. of Readings that can be stored in
Logging Mode
Max. Reading
Periodic
150000
Key Press
484
•
In Periodic mode, changing of any Measure or Source parameter is not allowed. So
While Periodic Logging is Running, the úser can’t enter
MEASURE
,
SOURCE
and
DISPLAY
menu. But in Key Press Logging mode, there is no restriction.
•
In Periodic Mode, if error message like “
Not Sufficient Memory
” comes while starting
the Logging. Try to Reduce Logging Time or Increase Sampling Period or try deleting
some existing files.
•
In Key Press Mode, If No of Samples reach its maximum limit that is 484, the next
sample will start from the first overwriting the memory.
•
While Logging is running, entering into the
LOGGING
menu shows below page.