Parameter 
Name 

 

Description / Options 

 

Trigger 

 

Data Logging Trigger Mode Selection.  

  
Available Options:  

  

Key Press  

Log Data on pressing 

 from RUN key Page  

Periodic  

Log Data periodically at every Sampling Rate for 

total time specified by Logging Time. 

 

Mode 

 

Data Mode Selection for Logging  

  
Available Options:  

  

EM  

Log only EM Measure Readings 

TC  

Log only TC Terminal Readings 

EM+TC  

Log EM Measure andTC Terminal both Readings 

  
This parameter is enabled only for Periodic Trigger.  

Save Method 

 

Reading Type selection for Logging  

  
Available Options:  

Instant  

Min  

Max  

Average  

All  

  
This parameter is enabled only for Periodic Trigger.  

Sampling 
Rate(s) 

 

Sampling Rate for Periodic Data Logging in seconds 

  
Range:  

1 … 9999  

  
This parameter is enabled only for Periodic Trigger.  

Logging Time 
(HH:MM:SS)

 

Total Logging Time in HH:MM:SS Format for Periodic Logging 

  
This parameter is enabled only for Periodic Trigger.  

Summary of Contents for PCE-TTC 30

Page 1: ...mperature Sensor User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last chang...

Page 2: ...ery 8 3 Start Up and Basic Operations 9 3 1 Power On or Off 9 3 2 User Interface 9 4 Menu Layout 20 4 1 MENU page 20 4 2 MEASURE Page 21 4 3 SOURCE Page 23 4 4 DISPLAY Page 30 4 5 DATA LOGGING Page 34...

Page 3: ...ible damage If any damage is visible do not use the device Do not use the instrument in explosive atmospheres The measurement range as stated in the specifications must not be exceeded under any circu...

Page 4: ...ent does not operate according to specifications contact the purchasing sales office as soon as possible Delivery contents 1 x Temperature sensor PCE TTC 30 2 x Laboratory cables with 2 mm plug 1 x Th...

Page 5: ...the next picture as well as all pictures of the PCE TTC 30 in this manual has an example configuration of modules The configuration of your PCE TTC 30 may vary significantly from the one in the pictur...

Page 6: ...TC 30 as the loop power supply while at the same time measuring the current or simply measuring the current while an external power supply is used The following picture displays the connection for Cur...

Page 7: ...hermocouples and mV This terminal accepts a miniature polarized thermocouple plug with flat inline blades spaced 7 9 mm 0 312 in center to center TC Terminal Measure and Source TC Type Range Display R...

Page 8: ...ies in different menus These are shown in the bottom left part of the display Scroll down to next parameter Decrease value of digit in Editbox Scroll down to previous parameter Increase value of digit...

Page 9: ...upplied with the device is USB A type male to USB B type male It is used for connecting charger and PC 2 2 5 Stand for Table Top Use This stand offers the best support for table top use which gives go...

Page 10: ...e Charging method Charging time to full capacity External Charging Adaptor 5 hours Note USB mini Type B connector charges the battery when the instrument is off and increases the battery life when the...

Page 11: ...e instrument shows a Startup Message and then shows the applicable data To set the instrument power off press and hold 2 seconds this button again When the power is off the last set of configuration o...

Page 12: ...to get source feedback reading 5 Data log memory corrupt 6 Source mode not working 9 More than one error from above list is occurring 3 USB Connection Status Icon Icon is visible if USB charging adapt...

Page 13: ...bar at the bottom of the display is visible all the time 2 function keys are available The meaning of the function keys varies depending on the situation A blank function key text means that the func...

Page 14: ...nit is Fahrenheit K Shows If TC Display mode is Actual and TC Unit is Kelvin mV Shows If TC Display mode is Actual And TC Type is 10 80 or 10 250mV Shows If TC Display mode is Percentage STEP RAM P Ic...

Page 15: ...Display Mode Actual Displays the Raw Input Value without any scaling Percentage Displays the Percentage Value Scaled Displays the Scaled Value Bar Graph Horizontal Bar graph according to TC Percentag...

Page 16: ...to Input Value Tare The Tare Value Set from MENU DISPLAY EM Terminal Tare page Actual Value The Raw Input Value without any scaling This will appear only if Main Display in MENU DISPLAY EM Terminal is...

Page 17: ...g Additional Info Unit Refer to TC Display Mode on previous pages EM Input Type EM Reading HART Icon Refer to EM Display Mode on previous pages EM Display Mode The Measure Reading Display Mode Blank D...

Page 18: ...tBox ListBox are used when there is a limited amount of preset values You have to select one of the available options The list of available options is displayed in the centre part A ListBox list opens...

Page 19: ...an be incremented or decremented The modified value can be saved using MENU ENT key The above figure shows the example how to change Input High 100 Range from 24 000 to 10 000 mA There are mainly 2 ty...

Page 20: ...PCE Instruments 18 To change the sign of the value shift to the sign digit and pressing UP or DOWN key will toggle the sign...

Page 21: ...Box is used where very few values can be possible and all the available values need to be visible In this device two types of Radio Button Box are available One with 1 value can be selectable the othe...

Page 22: ...easure Mode like Input Type Range etc TC SETUP Contains Parameters related to Thermocouple like TC Mode TC Type etc DISPLAY Contains Parameters related to different display mode for RUN page LOGGING C...

Page 23: ...Type Input Type Measure Input Type Available Options mA 0 000 24 000 mA DC mA 24 V 0 000 24 000 mA DC V 0 000 30 000 V DC Input Range Low 0 Low Range for Measure Input Range Default Input Low to Input...

Page 24: ...SETUP is set to Scaled Scaled Input Range High 100 Scaling High Range for Measure Input Range Scaled Input Range Low 0 to 99999 Decimal Point for this EditBox is changeable This parameter is enabled i...

Page 25: ...Page This page appears when you select RUN MENU TC SETUP This page contains parameters related to Thermocouple like TC Mode Type TC Type Unit TC Source Mode etc The Description of the Parameters appe...

Page 26: ...imulation Available Options TC Type Range Resolutio E TC 200 0 1000 0 C 0 1 C J TC 200 0 1200 0 C K TC 200 0 1372 0 C T TC 200 0 400 0 C B TC 450 0 1800 0 C R TC 0 0 1750 0 C S TC 0 0 1750 0 C N TC 20...

Page 27: ...Mode is MEASURE 4 3 1 STEP Page Parameter Name Description Options Low Starting Value of Step Enter value according to TC Display Mode If display mode is actual enter value in temperature mV and if d...

Page 28: ...unit Only appears if Step Definition is Temperature or Percentage Define Steps User Defined Step value for Manual and Auto Step Mode This option appears only if Step Definition is User Defined Maximu...

Page 29: ...TING Auto Stepping To Enable Auto Stepping select Source Type as STEP UnCheck the Manual Check Box If this mode is enabled Step UP or Step Down icon will appear in Source Display Window in RUN Page an...

Page 30: ...f Ramp Enter value according to TC Display Mode If display mode is actual enter value in temperature mV and if display mode is enter value in Hold 0 s Time to wait at Low 0 level in second This parame...

Page 31: ...Time s Time to decrease from High to Low Level Range 1 9999 Repeat Format How the Ramp should be done Available Options UP DOWN UP DOWN DOWN UP Repeat Repeat Counts Defines how many times the steps a...

Page 32: ...pressing F1 and F2 key running RAMP can be PAUSE or STOP at any time in RUN Page NOTE While RAMP is running RAMP settings can t be accessible and Source Page Parameter settings can t be change Stop RA...

Page 33: ...ay the Actual Input Value Percentage Display the Percentage Value of the Input The Value depends on Input Range These settings are available from MENU EM SETUP Scaled Display the Scaled Value of the I...

Page 34: ...easurement value 4 4 2 TC Display Settings This page appears when you select RUN MENU DISPLAY TC Terminal Parameter Name Description Options Main Display Select which Reading to be display as a Main R...

Page 35: ...erature with adding CJ Temperature mV Maximum Shows the Maximum measured reading from the time when info last reset Minimum Shows the Minimum measured reading from the time when info last reset Min Ma...

Page 36: ...Reading mV w o CJC Shows the Thermovoltage according to Temperature for CJ Temperature 0 C Filter sec 1st Order IIR Low Pass Filter for TC Measure Reading This option is available only for TC mode is...

Page 37: ...s TC Log only TC Terminal Readings EM TC Log EM Measure andTC Terminal both Readings This parameter is enabled only for Periodic Trigger Save Method Reading Type selection for Logging Available Option...

Page 38: ...0000 Key Press 484 In Periodic mode changing of any Measure or Source parameter is not allowed So While Periodic Logging is Running the ser can t enter MEASURE SOURCE and DISPLAY menu But in Key Press...

Page 39: ...opped when defined log time ends and a message LOGGING DONE pop ups Transferring the Results to a Personal Computer A 32 bit Windows software called mCAL exe can be downloaded here https www pce instr...

Page 40: ...f TC Terminal temperature CJC Adjustment CJ Temperature adjustment for Manual CJC Temperature 4 7 SETTING Page This page appears when you select RUN MENU SETTING All the available Settings Options are...

Page 41: ...rate HART devices This option is applicable for mA 24V Read Power Input Type Only 4 7 2 Display Settings Display Intensity Display Brightness Settings Range 5 to 100 Display Off Time Standby Time in s...

Page 42: ...ts re calibration may be necessary in some cases For calibrating the instrument a reliable source is required This source should be at least ten times accurate compared to the range of the instrument...

Page 43: ...s Error Code on status bar One of the peripheral not working properly Solution Restart the Device if still error code showing contact PCE Instruments Calibration Out Distortion in due to external nois...

Page 44: ...F 0 3 C 0 54 F J 200 1200 C 328 2192 F 0 1 C 0 18 F 0 3 C 0 54 F K 200 1372 C 328 2501 F 0 1 C 0 18 F 0 3 C 0 54 F T 200 400 C 328 752 F 0 1 C 0 18 F 0 3 C 0 54 F B 450 1800 C 842 3272 F 0 1 C 0 18 F...

Page 45: ...d Output voltage current loop 24V DC 24mA HART mA loop resistance 250 20 Special features Step and ramp function Automatic and manual mode x x2 For the measuring function Continuity test Adjustable th...

Page 46: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Page 47: ...3 975 114 Fax 1 561 320 9176 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce americas com info industrial needs com info pce italia it www pce instruments com us www pce instruments com english www...

Reviews: