CALIBRATION POINT REFERENCE 

Model 

psi-range 

Calibration point 

(±) Recommend 

PCE-P05 

0 ~ ±5 

4,000 

3,900 ~ 4,100 

PCE-P15 

0 ~ ±15 

12,00 

11,70 ~ 12,30 

PCE-P30 

0 ~ ±30 

24,00 

23,40 ~ 24,60 

PCE-P50 

0 ~ ±100 

80,00 

78,00 ~ 82,00 

MANUAL ZERO SETTING 

When you set the display to zero (no pressure applied to the connector), press button 

 for 2 seconds, now 

the meter display ''0 .0.0.0'' from right to left and then disappear each ''0'' from left back to right, the LCD display 

shows a normal mode. 

TROUBLESHOOTING 

  Power on but no display. Check the batteries are in place and making good contact or correct polarity, 

replace a new battery or attach optional AC adaptor for the weak battery caused. 

  BAT indication. Replace with a now battery when LCD display BAT at the middle bottom. 

  No Display. Make sure battery is not empty, if the display disappears, check sleep mode is active. Refer 

to  the  Disable  sleep  mode  function  for  a  long  time  using  the  measurement.  Or  check  the  tubing  is 

connected to the meter tightly. 

  Err.1.  For  the  pressure  value  exceed  the  maximum  range,  ''Err.  1''  appears  on  the  display,  p1ease 

change the sensor, otherwise, the sensor will be hurt for going on the over range measurement. 

  Err.2. For the measurement pressure is less than minimum range, ''Err. 2'' will appear recommend to 

change the sensor (meter). 

  Err.3.  For  operating  the  DIF  function,  the  differential  pressure  value  is  larger  than  maximum  display 

digit, Err.3 appears on the display. 

  Err.4.  When  you  set  zero,  make  sure  you  have  disconnected  the  tubing,  no  pressure  applied  to  the 

connector. Then if you see an Err.4 appears on the display, it means the sensor or meter damaged. 

Return the unit to the store you purchased for repaired. 

Err.4 will be also appeared while the tube or hose is connecting during setting zero mode. 

  E1OL or E2UL. When you see the errors while operating RS232 software, it means pressure source is 

less or over than the range of the instrument. 

REPLACING THE BATTERY 

Replace your 9-volt battery when:  

  The BAT icon appears on the right of the screen. 

  The meter will not power on. 

  Use of the back-light causes the BAT icon to appear. 

Even if the battery was recently replaced check its voltage level if you get no response from your instrument. 

To replace the battery:  

1.  Remove the tubing of the instrument. 

2.  Lay the instrument face-down on a clean, flat surface. 

3.  Remove the battery by screw driver and observe indicated polarity and close the cover after replacing 

with a new battery. 

 

Remove battery from instruments that you do not plan to use for a month or more. 

Do not leave battery in instrument. 

OPERATING CONDITIONS 

  Compensated temperature range: 0 – 50°C  

  Operating temperature 0 – 50°C (32 – 122°F)  

  Storage temperature range: -20 – 60°C 

  Operating Humidity Max. 80% RH 

  Power: One 9.0 volt battery 

  Exceeding Maximum pressure will cause permanent sensor damage. 

 

 

Summary of Contents for PCE-P05

Page 1: ...Instruction Manual HVAC Meter PCE PXX ...

Page 2: ...TO POWER OFF SLEEP FUNCTION 4 MODE OPTIONS 4 MAINTENANCE 5 CALIBRATION MODE 6 CALIBRATION POINT REFERENCE 7 MANUAL ZERO SETTING 7 TROUBLESHOOTING 7 REPLACING THE BATTERY 7 OPERATING CONDITIONS 7 SPECIFICATION 8 RS232 OUTPUT 8 WARRANTY 8 RETURN AUTHORIZATION 8 ...

Page 3: ...essure in psi that is referenced to ambient pressure and Differential pressure a measure of the difference two pressures 11 pressure units are selectable for Imperial and Metric in the different area bar mmHg ozin kgcm psi inH2 kPa ftH2O inHg cmH2O mbar Please check the tubing is not leakage or damaged before using CONVERSION Manometer PSI Inch of H2O mbar kg cm PCE P05 5 138 345 0 35 PCE P15 15 4...

Page 4: ... the modes on Default s y and relative in the middl y again and ord mode is d r 3 seconds t etal lug 5mm chasing the u TION cally in appr ger periods o he screen at ter paper off ble user sele ast used ed during the s of operatio setting and o e clock starts le deft of Fi the unit cycl displayed on to turn off the m plastic lug unit rox 20 minut of time you t which tinge ctable start u e last oper...

Page 5: ...t to the f zero before m clean the cas age the unit s e reading F op of the LC splay to show he unit return ce between t process con cle through the bottom o minates for a gure when ca shipping To ong time usin will maintain factory or a measurement se of the Ma s case with a Fig E CD and tile w as 0 0 o ns to the norm two pressure nection ope bar mmHg of the display approx 30 se alibration is b o...

Page 6: ...s nother 2 seco i e no SA b if you en ation when e or model 82 ative pressur d Manomete zero no pres appears on ure unit to b ation point t to increas tion point as key SA e negative p r the negativ all CA Se sing ke onds the me appeared p nter the righ entered to the 2100 i e th re er calibrator o ssure applied n the display be pointed u he adjustabl e the figure required and small ressure p e pr...

Page 7: ...DIF function the differential pressure value is larger than maximum display digit Err 3 appears on the display Err 4 When you set zero make sure you have disconnected the tubing no pressure applied to the connector Then if you see an Err 4 appears on the display it means the sensor or meter damaged Return the unit to the store you purchased for repaired Err 4 will be also appeared while the tube o...

Page 8: ...he D sub 9P type connector onto computer s Com 1 or 2 port or 3 Start to set up RS232 software by inserting the CD ROM or Floppy diskette 4 When installing the RS232 software please follow the operation manual procedure in the software package WARRANTY The meter is warranted to be free from defects in material and workmanship for a period of one year from the date of purchase This warranty covers ...

Reviews: