49
wattage should be multiplied by the VT ratio.
For example,
±0.2VAR becomes ± 0.2VAR * VT ratio
Range of VT (Voltage Transformer) Ratio: 1 to 3000
AC Active Energy (WH, or KWH, from 0 WH to 999,999 KWH)
WH = W * Time (in hours)
AC Current
(50 or 60 Hz, Auto Range, True RMS, Crest Factor < 4, Overload
Protection AC 2000A)
Range
Resolution
Accuracy of Readings
4.0 – 1500.0 A
0.01 A
±1.0% ± 5dgts
AC Voltage
(50 or 60 Hz, Auto Range, True RMS, Crest Factor < 4, VT=1, Input
Impedance 10 M
Ω
, Overload Protection AC 800V)
Range
Resolution
Accuracy of Readings
1
4.0 V – 600.0 V
0.1 V
±0.5% ± 5dgts
1
For VT
≠≠≠≠
1, the accuracy in percentage is the same (±0.5%). But the
additional digits should be multiplied by the VT ratio.
For example,
±5digits becomes ± 5digits * VT ratio
Harmonics of AC Voltage in Percentage
( 1 to 50
th
order, minimum voltage at 50
or 60 Hz > AC 80V. If the voltage is 0 at 50 or 60 Hz, all the percentage (%) display
is 0.)
Range
Resolution
Accuracy
1 – 20
th
0.1%
±2%
21 – 50
th
0.1%
4% of reading ±2.0%
Harmonics of AC Voltage in Magnitude
(1 to 50
th
order, minimum voltage at 50 or
60 Hz > AC 80V, VT=1)
Range
Resolution
Accuracy
1 – 20
th
0.1%
±2% ± 0.5V
21 – 50
th
0.1%
4% of reading ±0.5V
Summary of Contents for PCE-GPA 62
Page 1: ...Graphic AC Power Quality Analyzer PCE GPA 62 Users Manual PCE Group...
Page 11: ...6 HOUR MINUTE STOP TIME schedule when to stop datalogging HOUR...
Page 14: ...9 TOLERANCE...
Page 17: ...12 MONTH DAY...
Page 20: ...15 NOTE Frequency Hz is displayed in the second page Harmonics order 25 th to 50 th order...
Page 23: ...18 NOTE Frequency Hz is displayed in the second page Harmonics order 25 th to 50 th order...
Page 24: ...19 III 4 Waveforms of ACV andACAwith Phase Angle...
Page 30: ...25 III 5 6 RMS values of Voltage and Current...
Page 35: ...30...
Page 45: ...40 Press the FUNC button to start transient detection...
Page 48: ...43...
Page 58: ...53...