48
XI. SPECIFICATIONS (23
°°°°
C±5
°°°°
C)
AC Watt (50 or 60 Hz, PF 0.5 to 1.0, VT = 1, Voltage > 5V AC, Current > 5A AC
for A range, and continuous waveform)
Accuracy of Readings
1
Range (0 to 1500A)
Resolution
> 20 V and > 20A
< 20V or < 20A
10.0 – 999.9 W
0.1W
±1% ±20dgts
±2% ±40dgts
1.000 – 9.999 KW
0.001 KW
±1% ±20dgts
±2% ±40dgts
10.00 – 99.99 KW
0.01 KW
±1% ±20dgts
±2% ±40dgts
100.0 – 999.9 KW
0.1 KW
±1% ±20dgts
±2% ±40dgts
1000 – 9999 KW
1 KW
±1% ±20dgts
±2% ±40dgts
1
For VT
≠≠≠≠
1, the accuracy in percentage is the same (±1%). But the additional
wattage should be multiplied by the VT ratio.
For example,
±0.2W becomes ±0.2W * VT ratio
AC Apparent Power (VA, from 0.000VA to 9999 KVA, PF 0.5 to 1)
Accuracy of Readings
1
Range (0 to 1500A)
Resolution
> 20 V and > 20A
< 20V or < 20A
10.0 – 999.9 VA
0.1VA
±1% ±20dgts
±2% ±40dgts
1.000 – 9.999 KVA
0.001 KVA
±1% ±20dgts
±2% ±40dgts
10.00 – 99.99 KVA
0.01 KVA
±1% ±20dgts
±2% ±40dgts
100.0 – 999.9 KVA
0.1 KVA
±1% ±20dgts
±2% ±40dgts
1000 – 9999 KVA
1 KVA
±1% ±20dgts
±2% ±40dgts
1
For VT
≠≠≠≠
1, the accuracy in percentage is the same (±1%). But the additional
wattage should be multiplied by the VT ratio.
For example,
±0.2VA becomes ±0.2VA * VT ratio
AC Reactive Power (VAR, from 0.000 VAR to 9999 KVAR)
Accuracy of Readings
1
Range (0 to 1500A)
Resolution
> 20 V and > 20A
< 20V or < 20A
10.0 – 999.9 VAR
0.1VAR
±2% ±30dgts
±3% ±40dgts
1.000 – 9.999 KVAR
0.001 KVAR
±2% ±30dgts
±3% ±40dgts
10.00 – 99.99 KVAR
0.01 KVAR
±2% ±30dgts
±3% ±40dgts
100.0 – 999.9 KVAR
0.1 KVAR
±2% ±30dgts
±3% ±40dgts
1000 – 9999 KVAR
1 KVAR
±2% ±30dgts
±3% ±40dgts
1
For VT
≠≠≠≠
1, the accuracy in percentage is the same (±2%). But the additional
Summary of Contents for PCE-GPA 62
Page 1: ...Graphic AC Power Quality Analyzer PCE GPA 62 Users Manual PCE Group...
Page 11: ...6 HOUR MINUTE STOP TIME schedule when to stop datalogging HOUR...
Page 14: ...9 TOLERANCE...
Page 17: ...12 MONTH DAY...
Page 20: ...15 NOTE Frequency Hz is displayed in the second page Harmonics order 25 th to 50 th order...
Page 23: ...18 NOTE Frequency Hz is displayed in the second page Harmonics order 25 th to 50 th order...
Page 24: ...19 III 4 Waveforms of ACV andACAwith Phase Angle...
Page 30: ...25 III 5 6 RMS values of Voltage and Current...
Page 35: ...30...
Page 45: ...40 Press the FUNC button to start transient detection...
Page 48: ...43...
Page 58: ...53...